Online: 6
Today: 94
Month: 104479
Total Access: 2603715
Range
+ FNU Mode: 0.00 to 9.99; 10.0 to 99.9; 100 to 1000 FNU
+ FAU Mode: 10.0 to 99.9; 100 to 4000 FAU
+ NTU Ratio Mode: 0.00 to 9.99; 10.0 to 99.9; 100 to 4000 NTU 0.00 to 9.99; 10.0 to 99.9; 100 to 980 EBC
+ NTU Non-Ratio Mode: 0.00 to 9.99; 10.0 to 99.9; 100 to 1000 NTU 0.00 to 9.99; 10.0 to 99.9; 100 to 245 EBC
Resolution
+ FNU Mode: 0.01; 0.1; 1 FNU
+ FAU Mode: 0.1; 1 FAU
+ NTU Ratio Mode: 0.01; 0.1; 1 NTU / 0.01; 0.1; 1 EBC
+ NTU Non-Ratio Mode: 0.01; 0.1; 1 NTU / 0.01; 0.1; 1 EBC
Accuracy @25°C
+ FNU Mode: ±2% of reading plus stray light
+ FAU Mode: ±10% of reading
+ NTU Ratio Mode: ±2% of reading plus stray light / ±5% of reading above 1000 NTU
+ NTU Non-Ratio Mode: ±2% of reading plus stray light
Range Selection: automatic
Repeatability: ±1% of reading or stray light, whichever is greater
Stray Light: < 0.1 NTU (0.05 EBC)
Light Detector: silicon photocell
Light Source: IR LED
Method: ISO 7027 Method
Measuring Mode: normal, average, continuous.
Turbidity Standards: <0.1, 15, 100, 750 FNU and 2000 NTU
Calibration: two, three, four or five-point calibration
Log Memory: 200 records
PC Connectivity: USB
Environment: 0°C (32°F) to 50°C (122°F); max 95% RH non-condensing
Power Supply: 12 Vdc power input
Dimensions / Weight: 230 x 200 x 145 mm (9 x 7.9 x 5.7”) / 2.5 Kg (88 oz.)
Order Information: HI 88713-02 (230V) are supplied with sample cuvettes and caps (6), calibration cuvettes, silicone oil, tissue for wiping cuvettes, power adapter and instruction manual.
Warranty: 12 Months